优质产品

专业水平

诚实信誉

首页    思拓凡 Cytiva    蛋白纯化系统    ÄKTA avant 全自动智能层析系统

产品特性

ÄKTA avant 是一种制备级蛋白纯化层析系统,设计用于快速、安全地开发可扩大性蛋白方法和工艺。ÄKTA avant是用于方法探索和纯化工艺优化的全自动智能液相色谱系统,现有两个型号。ÄKTA avant两款系统具有不同的流速/压力参数,但具有相同的硬件装置。ÄKTA avant 25系统设计流速为25mL/min,用于填料筛选,方法优化和数百毫克级样品分析和制备。ÄKTA avant 150系统设计流速为150mL/min,可用于填料筛选方法优化和工艺放大,以及克级样品的分析和制备。

ÄKTA avant 全自动智能层析系统

产品详情

ÄKTA avant 25

参数 ÄKTA avant 25
Application Fast and secure process development and method optimization 
Application supported Affinity chromatography, gel filtration/size exclusion chromatography, ion exchange chromatography, and hydrophobic interaction chromatography 
Automatic buffer preparation BufferPro 
Autosampler injection No 
Buffer selection 18 buffer inlets, option of up to 46 
Compliance The product fulfills valid directives and standards when used within the conditions specified in the user manual. The product must also be used in the same state as it was delivered from Cytiva and connected only to other CE labeled Cytiva modules or other products as recommended. For regulatory details please see declaration of conformity. 
Depth 710 mm 
Flow cell path length 2 mm (optional 10 mm) 
Flow rate 0.001 mL/min-25 mL/min 
Height 660 mm 
Includes Accessory box, 3 mixer chambers (0.6, 1.4 and 5 mL), pH electrode, fraction collector cassette tray, 4 deep-well cassettes, 2 cassettes for 50mL tubes 
Multiple sample injection Yes, up to 7 samples 
Multiple wavelength detection Yes, up to 3 wavelengths 
Multistep purification Yes (optional) 
Operating pressure (sample pump) max. 10 MPa (100 bar, 1450 psi) 
Operating pressure max. 20 MPa (200 bar, 2900 psi) 
Power consumption 800 VA 
Protection class IP 21, wet side IP 22 
Pump type Piston pump, metering type 
Relative humidity 20% to 95% (non condensing) 
Safety certifications 2006/42/EC (MD), 2004/108/EC (EMC), 1999/5/EC (R&TTE) 
Tubing i.d. (flow path) 0.75 mm 
UV wavelength 190–700 nm 
Viscosity 0.35–10 cP (system pump), 0.7–10 cP (sample pump) 
Voltage 100–120/220–240 VAC 
Width 860 mm
pH monitoring 0–14 
Model ÄKTA avant 25 

 

ÄKTA avant 150

参数 ÄKTA avant 150
Application Fast and secure process development and scale-up 
Application supported Affinity chromatography, gel filtration/size exclusion chromatography, ion exchange chromatography, and hydrophobic interaction chromatography 
Automatic buffer preparation BufferPro 
Autosampler injection No 
Buffer selection 18 buffer inlets, option of up to 46 
Compliance The product fulfills valid directives and standards when used within the conditions specified in the user manual. The product must also be used in the same state as it was delivered from Cytiva and connected only to other CE labeled Cytiva modules or other products as recommended. For regulatory details please see declaration of conformity. 
Depth 710 mm 
Flow cell path length 2 mm (optional 10 mm and 0.5 mm) 
Flow rate 0.01 mL/min-150 mL/min 
Height 660 mm 
Includes Accessory box, 3 mixer chambers (1.4, 5 and 15 mL), pH electrode, fraction collector cassette tray, 2 cassettes for 15 mL tubes, 4 cassettes for 50 mL tubes. 
Multiple sample injection Yes, up to 7 samples 
Multiple wavelength detection Yes, up to 3 wavelengths 
Multistep purification No 
Operating pressure (sample pump) max. 5 MPa (50 bar, 725 psi) 
Operating pressure max. 5 MPa (50 bar, 725 psi) 
Power consumption 800 VA 
Protection class IP 20 
Pump type Piston pump, metering type 
Relative humidity 20% to 95% (noncondensing) 
Safety certifications 2006/42/EC (MD), 2004/108/EC (EMC), 1999/5/EC (R&TTE) 
Tubing i.d. (flow path) 0.75 mm 
UV wavelength 190–700 nm 
Viscosity 0.35 to 5 cP (system pump), 0.7 to 10 cP (sample pump) 
Voltage 100–120/220–240 VAC 
Width 860 mm
pH monitoring 0–14 
Model ÄKTA avant 150